Performance specifications
OP mesh trays: 10
AN sets: 4
MI sets: 2
Narrow-necked laboratory glassware: 72 per program
Control system
Freely programmable, electronic Profitronic controls
4-line, full-text navigation display of program, program sequence, temperatures and faults
64 program slots
8 service programs and
9 standard programs ex works
47 additional customisable slots
Rotary program selector switch
Program compilation direct or via a PC/laptop
Appropriate codes can be entered to restrict program access and programming of system parameters
RS 232 serial interface
9 potential-free output signals
3 potential-free input signals
Features
2 Qmax circulation pump 300/400 l/min
AV Drain valve
AP Drain pump;
TA 2 Hot-air drying units
BO Boiler for pre-heating demineralised water
Heating
Steam, electric; steam/electric (convertible)
Dispensing technology
2 dispenser pumps/hose
Cleaning technology
Fresh water system
Thermal disinfection at 93°C with 10 min. holding time in accordance with § 18 IfsG (German legislation on the containment of epidemics) to ensure fungicidal, bactericidal and virucidal disinfection
Patented DESIN vario TD process for rigid fibre optics, drive systems, patient systems.
Baskets and inserts are directly coupled to the water circulation and hot-air drying system
Design
Frontloading unit with bottom-hinged door
Outer casing: AE stainless-steel
Large cabinet
Stainless steel cabinet and spray arms
2-skin construction:
H 1975, W 900, D 750 mm
(H incl. SBW/2 plinth/drip tray)
H 2595, W 900, D 750 mm
(H incl. SBW/2 plinth/drip tray and MAV/2 top panel installation kit for DK/2 drying unit and/or steam condenser)
Test certificates
EMV/Interference suppression; DVGW; VDE; CE 0366; § 10c BSeuchG